dc.description.abstractIn this study we present an NIR carbon dioxide gas sensor based on an inner filter process that includes an ionic liquid (IL), 1-ethyl-3-methylimidazolium tetrafluoroborate (EMIMBF4), to improve its stability, dynamic behavior and lifetime, which are usually the main drawbacks with these sensors. The presence of CO2 causes a displacement of a simple boron-dipyrromethene-type fluorophore, azaBODIPY, as the pH indicator towards its acid form. This increases the emission intensity of Cr(III)-doped gadolinium aluminium borate (GAB) as the luminophore. The characterization of the prepared sensor was carried out and a discussion of the results is presented. The response and recovery times improved considerably, 23 and 49 s, respectively, with respect to the sensor without IL, at 60 and 120 s, respectively,. Additionally, the measurement range is extended when using IL, able in this case to measure in the complete range up to 100% CO2; without IL the measurement range is limited to 60% CO2. The detection limit ranges from 0.57% CO2 without IL to 0.26% CO2 when IL is added. The useful lifetime of the sensing membrane was 20 days for membranes with IL and only 6 days for membranes without IL, with the sensor always kept in the dark and without the need to maintain a special atmosphere.es_ES
dc.language.isoenges_ES
dc.publisherMdpies_ES
dc.rightsAtribución 3.0 España*
dc.rights.urihttp://creativecommons.org/licenses/by/3.0/es/*
dc.subjectNIRes_ES
dc.subjectIonic liquidses_ES
dc.subjectCarbon dioxide determinationes_ES
dc.subjectFluorescent sensorses_ES
dc.subjectazaBODIPYes_ES
dc.titleNear Infrared Sensor to Determine Carbon Dioxide Gas Based on Ionic Liquides_ES
dc.typejournal articlees_ES
dc.rights.accessRightsopen accesses_ES
dc.identifier.doi10.3390/coatings11020163
dc.type.hasVersionVoRes_ES


Ficheros en el ítem

[PDF]

Este ítem aparece en la(s) siguiente(s) colección(ones)

Mostrar el registro sencillo del ítem

Atribución 3.0 España
Excepto si se señala otra cosa, la licencia del ítem se describe como Atribución 3.0 España