Smart quick response code for multianalyte determination
Metadatos
Mostrar el registro completo del ítemAutor
Marín-Sánchez, Antonio; Ortiz Gómez, Inmaculada; Gallego-Méndez, David; Rodríguez Domínguez, Carlos; Capitán Vallvey, Luis Fermín; Morales Santos, Diego Pedro; Álvarez-Bermejo, José Antonio; Castillo Morales, María Encarnación; Salinas Castillo, AlfonsoEditorial
Elsevier
Materia
Colorimetric Glucose Lactate
Fecha
2022-11-30Referencia bibliográfica
Published version: Marín-Sánchez, Antonio et al. Smart quick response code for multianalyte determination. Measurement Volume 204, 30 November 2022, 112035. https://doi.org/10.1016/j.measurement.2022.112035
Patrocinador
Spanish Ministerio de Economía y Competitividad (Projects PID2019-103938RB-I00, PID2020-117344RBI00 and CTQ2017-86125-P); Junta de Andalucía (Projects B-FQM-243-UGR18, P20_00265 and P18-RT-2961); European Regional Development Funds (ERDF)Resumen
The extensive use of smartphones with point-of-care applications has bolstered the
development of what has come to be known as Healthcare 4.0. The ability to track biomarker
parameters has become an essential part of monitoring the evolution of certain diseases, with
the potential for application with a large selection of population groups. In that vein, this
work presents a multianalyte quick response (QR) colorimetric code for the real-time
measurement of glucose and lactate samples using smartphones. The colorimetric QR code
contains both the QR static data and the dynamic information from the sample analyses. Two
correction methods are applied in order to compensate for the variation in the surrounding
conditions in the determination of the biomarkers. Glucose and lactate recovery between 91
and 117% was found for assays with sweat and serum samples using a smartphone readout.
The developed system demonstrates flexible measurement conditions for using in real
situations.