Evaluation of a total reflection x-ray fluorescene spectrometer in the determination of arsenic and trace metals in environmental samples
Identificadores
URI: https://hdl.handle.net/10481/96419Metadatos
Mostrar el registro completo del ítemMateria
TXRF total reflection X-ray fluorescence arsenic trace metals mine tailings surface water acid mine drainage
Fecha
2009Referencia bibliográfica
Juvonen, R., Parviainen, A., & Loukola-Ruskeeniemi, K. (2009). Evaluation of a total reflection x-ray fluorescene spectrometer in the determination of arsenic and trace metals in environmental samples. Geochemistry: Exploration, Environment, Analysis, 9(2), 173-178. https://doi.org/10.1144/1467-7873/09-205
Resumen
Suitability of the PicoTAX total reflection X-ray fluorescence (TXRF) spectrometer was studied for determination of As and trace metals in the Haveri Au-Cu sulphide mine tailings leached with aqua regia and in water affected by acid
mine drainage. Results for 12 water and 16 tailings samples were compared with accredited methods, inductively coupled plasma-atomic emission spectrometry (ICP-AES) and inductively coupled plasma-mass spectrometry (ICP-MS). Certified
reference samples analysed were QC Metal LL from Eurofins A/S, and groundwater CRM 609 and lake sediment BCR-280R from the Institute for Reference Materials and Measurements (IRMM) of the European Commission. Approximate method
detection limits (MDL) for water samples, based on reference sample analyses, are in the range of 20 μg l-1 for As, Cr, Pb and Zn and 5–10 μg l-1 for Co, Cu, Ni, V and Mn. MDLs for aqua regia leach, calculated using reagent blank values, were
20 (mg kg-1) for As, Co, Cr, Cu, Mn, Ni, V, and Zn. Random error of determination varied from 12 to 20 %. Inter-element interferences include: As and Pb on each other, and Fe on Cr and Co. Using the Mo anode of the X-ray tube
excludes the determination of Cd, Mo, Nb, Tc and Zr. TXRF can be recommended merely as a preliminary technique for selecting samples.