Mostrar el registro sencillo del ítem

dc.contributor.authorTejada-Casado, Carmen
dc.contributor.authorHernández-Mesa, Maykel 
dc.contributor.authordel Olmo-Iruela, Monsalud
dc.contributor.authorGarcía-Campaña, Ana María 
dc.date.accessioned2025-01-27T09:17:21Z
dc.date.available2025-01-27T09:17:21Z
dc.date.issued2016-08-02
dc.identifier.urihttps://hdl.handle.net/10481/100434
dc.description.abstractA novel method for the analysis of benzimidazole residues in water samples by capillary electrochromatography-UV detection (290 nm), using laboratory-made packed columns is presented. Capillaries (25 cm packed length×75 µm inner diameter, 34 cm total length, 25.5 cm effective capillary length) were packed with C18 particles (5 µm, non-encapped) following a high pressure packing procedure and using a compact steel unit designed for packing capillary columns. Acetone was employed as solvent to carry the particles through the capillary and pack it under a pressure of 42 MPa. Outlet and inlet frits were made by sintering the particles of the stationary phase by heating the packed material with a nichrome ribbon connected to a 7 V AC power supply. With the aim of achieving a good analytical performance, the variables that affected the separation were studied, using a mobile phase composition of 60:40 (v/v) acetonitrile/water containing ammonium acetate (1 mM, pH 6.5), a separation voltage of 25 kV and a temperature of 25 °C. In addition, a combined hydrodynamic-electrokinetic injection mode was considered and samples were injected for 75 s under a voltage of 12.5 kV and a pressure of 11.5 bar. Finally, the determination of benzimidazoles in water samples was accomplished by capillary electrochromatography using dispersive liquid-liquid microextraction as sample treatment. Variables affecting the extraction efficiency were optimized, using chloroform and ethanol as extraction and disperser solvents, respectively. This method was applied to different water samples, obtaining satisfactory results in terms of linearity (R2≥0.990), repeatability (RSD≤1.2%), reproducibility (RSD≤2.2%) and trueness (R≥87.7%). Detection and quantification limits were lower than 2.8 µg L−1 and 9.3 µg L−1, respectively.es_ES
dc.description.sponsorshipThe authors gratefully acknowledge the financial support of the Andalusian Government (Junta de Andalucía) for supporting this work (Excellence Project Ref: P12-AGR-1647). Maykel Hernández-Mesa thanks to the Plan Propio of University of Granada and Carmen Tejada-Casado thanks to the Excellence Project Ref: P12-AGR-1647 for the predoctoral fellowships.es_ES
dc.language.isoenges_ES
dc.rightsAttribution-NonCommercial-NoDerivatives 4.0 Internacional*
dc.rights.urihttp://creativecommons.org/licenses/by-nc-nd/4.0/*
dc.subjectBenzimidazole compoundses_ES
dc.subjectCapillary electrochromatographyes_ES
dc.subjectDispersive liquid-liquid microextractiones_ES
dc.subjectPacked capillary columnses_ES
dc.subjectWater sampleses_ES
dc.titleCapillary electrochromatography coupled with dispersive liquid-liquid microextraction for the analysis of benzimidazole residues in water sampleses_ES
dc.typejournal articlees_ES
dc.rights.accessRightsopen accesses_ES
dc.identifier.doi10.1016/j.talanta.2016.08.012
dc.type.hasVersionAMes_ES


Ficheros en el ítem

[PDF]

Este ítem aparece en la(s) siguiente(s) colección(ones)

Mostrar el registro sencillo del ítem

Attribution-NonCommercial-NoDerivatives 4.0 Internacional
Excepto si se señala otra cosa, la licencia del ítem se describe como Attribution-NonCommercial-NoDerivatives 4.0 Internacional