Determination of 5-nitroimidazole residues in milk by capillary electrochromatography with packed C18 silica beds
Metadatos
Mostrar el registro completo del ítemAutor
Hernández Mesa, Maykel; Lara, Francisco J; Cruces Blanco, María Carmen; García Campaña, Ana MaríaMateria
Capillary electrochromatography Packed capillary columns Slurry pressure packing 5-Nitroimidazoles Milk
Fecha
2015-06-22Patrocinador
The authors gratefully acknowledge the financial support of the Andalusian Government (Junta de Andalucía) for supporting this work (Excellence Project Ref: P12-AGR-1647). Maykel Hernández-Mesa thanks to Plan Propio of University of Granada for a pre-doctoral fellowship.Resumen
This work presents a novel methodology for analysing 5-nitroimidazole residues in milk samples by capillary electrochromatography using lab-made packed columns, produced by carrying out a high pressure packing procedure using acetone as driving solvent and C18 silica uncapped particles (5 µm particle size) as packing material. Column frits resulted from sintering the proper stationary phase by heating the packed material for 20 s with a nichrome ribbon (80% Ni–20% Cr, 28 cm×2 mm×0.2 mm, electric resistance 1.3 Ω) connected to a 7 V AC power supply. Lab-made C18 silica packed capillaries (40 cmx50 µm i.d.) were employed for the determination of 5-nitroimidazole drugs. Milk samples were treated by a salting-out assisted liquid–liquid extraction followed by a solid phase extraction with Oasis®HLB cartridges prior to their injection. Samples were hydrodynamically injected into the column for 120 s at 11.5 bar. Afterwards eight 5-nitroimidazole compounds were separated in isocratic mode under an applied voltage of 27 kV and a temperature of 30 °C. The selected mobile phase consisted of a mixture 60:40 acetonitrile:ammonium acetate (2.5 mM, pH=5). Separation was monitored at 320 nm and it was performed in less than 15 min. The method was characterized in terms of linearity (R2≥0.993) and precision (repeatability, RSD≤12.2% and reproducibility, RSD≤14.5%), obtaining detection limits lower than 29 µg/L for all compounds under study.