Assessing the test-retest reliability of MyotonPRO for measuring achilles tendon stiffness
Metadatos
Afficher la notice complèteAuteur
Volesky, Krystof; Novak, Jan; Janek, Michael; Katolicky, Jakub; Tufano, James J.; Steffl, Michal; Courel Ibáñez, Javier; Vetrovsky, TomasEditorial
MDPI
Materia
Achilles tendon stiffness reliability
Date
2025-02-28Referencia bibliográfica
Volesky, K.; Novak, J.; Janek, M.; Katolicky, J.; Tufano, J.J.; Steffl, M.; Courel-Ibáñez, J.; Vetrovsky, T. Assessing the Test-Retest Reliability of MyotonPRO for Measuring Achilles Tendon Stiffness. J. Funct. Morphol. Kinesiol. 2025, 10, 83. https:// doi.org/10.3390/jfmk10010083
Résumé
Objectives: This study evaluates the test-retest reliability and inter-rater reliability of the MyotonPRO for measuring Achilles tendon stiffness at two standardized sites
over various time frames and settings. Methods: Eight healthy participants underwent
assessments by three raters over six visits. Tendon stiffness was measured at proximal
(mid-portion) and distal (insertional) regions of the Achilles tendon at various time frames
(10–15 s, 10–15 min, 24 h, and 14 days apart). Measurements included participant repositioning and two activity stimuli (daily living and sport). Reliability was calculated using
the intraclass correlation coefficient (ICC), its 95% confidence interval, coefficient of variation, standard error of measurement, and minimal detectable change. Results: Short-term
reliability (10–15 min) was excellent, with an ICC of 0.956 (0.929–0.974). Between days
reliability (24 h) was good, with an ICC of 0.889 (0.802–0.938). Between weeks reliability
(2 weeks) was good with an ICC of 0.886 (0.811–0.931). Short-term reliability with the simulation of activity of daily living was good, with an ICC of 0.917 (0.875–0.945). Short-term
reliability with the simulation of sport was good with an ICC of 0.933 (0.891–0.96). Between
days reliability with the simulation of sport was good, with an ICC of 0.920 (0.859–0.955).
Conclusions: When used in a standardized position, the MyotonPRO demonstrates reliable
repeated measurements of Achilles tendon stiffness. This protocol provides a foundation
for clinical research and rehabilitation by clarifying expected reliability across minutes,
days, and weeks, thus aiding clinicians and researchers in monitoring tendon adaptations
and making evidence-based decisions.