@misc{10481/89957, year = {2012}, month = {9}, url = {https://hdl.handle.net/10481/89957}, abstract = {Objective: To evaluate the performance of the Sysmex UF1000i automatic urine screening system in the quantitative culture of fast-growth aerobic/facultative bacteria. Methods and results: A standard procedure was used to recover fast-growth aerobic/facultative micro-organisms in 1225 samples, applying (Sysmex(®)) flow cytometry for parallel bacteria and leucocyte counts. According to the area under the receiver operating characteristic curve, the optimal cut-off values to detect bacteriuria >10(5) colony forming units (CFU) ml(-1) were 690/μl for bacteria and 38/μl for leucocytes (sensitivity, 92%; specificity, 65%; positive predictive value [PPV], 39%; and negative predictive value [NPV], 97%). The use of a single cut-off point of 150 bacteria μl(-1) to detect significant bacteriuria of >10(5) CFU ml(-1) or of ≥10(4) CFU ml(-1) plus leucocyturia obtained similar results (sensitivity, 89%; specificity, 54%; PPV, 31%; and NPV, 96%) and allowed 45.7% of the samples to be rapidly excluded. Conclusions: The Sysmex UF1000i system can be adapted for bacteriuria screening by the use of an appropriate cut-off point. Significance and impact of the study: This screening system significantly reduces the workload and produces very few false positives and negatives.}, publisher = {Oxford University Press}, keywords = {Bacteriuria}, keywords = {Cytometry}, keywords = {Sysmex}, keywords = {Urinary tract infection}, keywords = {Urine}, title = {Performance of the Sysmex UF1000i system in screening for significant bacteriuria before quantitative culture of aerobic/facultative fast-growth bacteria in a reference hospital}, doi = {10.1111/j.1365-2672.2012.05369.x}, author = {Gutiérrez Fernández, José and Lara, A. and Bautista, M.F. and Luna Del Castillo, Juan De Dios and Polo, P. and Miranda, C. and Navarro, J.M.}, }