@misc{10481/87249, year = {2007}, url = {https://hdl.handle.net/10481/87249}, abstract = {This paper presents the study of a disposable optical sensor to determine potassium by an ion-exchange mechanism based on the conventional ionophore dibenzo-18-crown-6. The normalised signal used as analytical parameter is obtained from measurements in transflectance mode and the results are compared with those obtained by transmission mode. The sensor responds linearly in activities up to 100mM with a detection limit of 19mM and a precision around 7%. It is possible to linearise the sigmoidal calibration curve by means of a decimal logistic transformation applied to transflectance data, in this way increasing the linear dynamic range with a detection limit of 8.4 mM. The potential utility of the disposable sensor was checked in the determination of potassium in different types ofwaters, validating the results against a reference procedure. The viability of the semi-quantitative determination of potassium using this disposable sensor was studied using a colour-difference CIELAB-base formula, designed to give quantitative values of the perceived colour difference between a pair of coloured samples under specific experimental conditions, obtaining seven intervals of discriminable concentrations in the colour chart between 10−7 and 10−1 M.}, organization = {We acknowledge financial support from the Ministerio de Educación y Cultura, Dirección General de Enseñanza Superior (Spain) (Projects CTQ2005-09060-CO2-01 and CTQ2005- 09060-CO2-02)}, publisher = {Elsevier}, title = {Potassium disposable optical sensor based on transflectance and cromaticity measurements}, doi = {10.1016/j.snb.2007.05.016}, author = {Orbe Payá, Ignacio De and Erenas Rodríguez, Miguel María and Capitán Vallvey, Luis Fermín}, }