@misc{10481/31427, year = {2009}, url = {http://hdl.handle.net/10481/31427}, abstract = {A commercial, position-sensitive ion detector was used for the first time for the time-of-flight ion-cyclotron resonance detection technique in Penning trap mass spectrometry. In this work, the characteristics of the detector and its implementation in a Penning trap mass spectrometer will be presented. In addition, simulations and experimental studies concerning the observation of ions ejected from a Penning trap are described. This will allow for a precise monitoring of the state of ion motion in the trap.}, organization = {The support of the Helmholtz Association of National Research Centres (HGF) under Contract number VH-NG-037, of the Deutsche Forschungsgemeinschaft (BL981-2-1), and of the German Federal Ministry for Education and Research (06MZ215) is gratefully acknowledged. We also acknowledge the support of Horst Schmidt-Böcking and the RoentDek company. Sz. Nagy acknowledges support from the Alliance Program of the Helmholtz Association (HA216/EMMI).}, publisher = {Elsevier}, keywords = {Delay-line detector}, keywords = {Penning trap mass spectrometry}, keywords = {TOF-ICR}, keywords = {Ion optics}, keywords = {Beam diagnostics}, keywords = {Position-sensitive ion detection}, title = {Position-sensitive ion detection in precision Penning trap mass spectrometry}, doi = {10.1016/j.nima.2009.04.046}, author = {Eitel, G. and Block, M. and Czasch, A. and Dworschak, M. and George, S. and Jagutzki, O. and Ketelaer, J. and Ketter, J. and Nagy, Sz. and Rodríguez, Daniel and Smorra, C. and Blaum, K.}, }